logo
logo

This document provides an overview of the specifications and factory default settings specific to your Proteus WeldSaverâ„¢ product. Please retain this sheet for future reference.

Additional product information is available on the Proteus Industries website at:


PRODUCT INFORMATION


Model Number

W6PVE150TCNC-NV1

Serial Number/MAC ID

00-23-92-00-CD-4F

Sale Order Number

144402

Product Series

WeldSaver Passport

Control Interface

Ethernet/IP

Firmware Revision

1.12.3


CONTROL PARAMETER SETTINGS



Outlet Flow Warning

11.4 LPM

Flow Fault

7.6 LPM

Slow Leak Difference

4 LPM

Slow Leak Delay

4000 ms

Stabilization Delay

4s

Leak Response

Normal


FLOW SENSOR CALIBRATION


Inlet Flow Sensor

Flowrate(LPM) Frequency(Hz)
15.12 77.71
24.79 131.27
39.84 206.43
39.84 206.43
39.84 206.43

Outlet Flow Sensor

Flowrate(LPM) Frequency(Hz)
15.12 78.18
24.79 130.69
39.84 209.97
39.84 209.97
39.84 209.97

NETWORK SETTINGS


IP Address

172.24.1.1

Netmask

255.255.255.0

Gateway

172.24.1.100


Prefered DNS

172.24.1.100

Alternate DNS

172.24.1.100

DHCP

Static


TEST RESULT


Flow Sensor Test

Passes

Functionality Test

Passes

Date

04/07/2025

Operator

ERNNY


Proteus Industries Inc. | 340 Pioneer Way, Mountain View, CA 94041 | Tel: (650) 964-4163 | Fax: (650) 965-0304

www.proteusind.com | sales@proteusind.com